| Model Number |
420172-18 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of customer provided image was performed by an intuitive surgical inc., (isi) regulatory post market surveillance analyst.Following information was provided : the image did not show visible damage(s) on the instrument cables.Intuitive surgical, inc.(isi) did receive the 8mm maryland bipolar forceps instrument to perform failure analysis.Failure analysis investigations confirmed the customer reported complaint.The instrument was analyzed and it was found to have a broken conductor wire within the bipolar yaw pulley, at the distal end.The instrument failed an electrical continuity test, confirming a complete break in the wire.No thermal damage was found on the instrument.The probable root cause is attributed to component susceptibility to damage through external collisions, during use, or during reprocessing.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|