MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; MARYLAND BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; MARYLAND BIPOLAR FORCEPS Back to Search Results
Model Number 420172-18
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
 
Event or Problem Description
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
 
Additional Manufacturer Narrative
A review of customer provided image was performed by an intuitive surgical inc., (isi) regulatory post market surveillance analyst.Following information was provided : the image did not show visible damage(s) on the instrument cables.Intuitive surgical, inc.(isi) did receive the 8mm maryland bipolar forceps instrument to perform failure analysis.Failure analysis investigations confirmed the customer reported complaint.The instrument was analyzed and it was found to have a broken conductor wire within the bipolar yaw pulley, at the distal end.The instrument failed an electrical continuity test, confirming a complete break in the wire.No thermal damage was found on the instrument.The probable root cause is attributed to component susceptibility to damage through external collisions, during use, or during reprocessing.
 
Event or Problem Description
Refer to h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
MARYLAND BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key22501847
Report Number2955842-2025-29610
Device Sequence Number8468188
Product Code NAY
UDI-Device Identifier00886874111475
UDI-Public(01)00886874111475(10)K10230223
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420172-18
Device Catalogue Number420172
Device Lot NumberK10230223 998
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2025
Supplement Date Received by Manufacturer09/15/2025
Initial Report FDA Received Date07/14/2025
Supplement Report FDA Received Date09/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-