| Model Number |
8888145014 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during dialysis treatment, it was found that the catheter had a crack at the blue (venous) luer connector.It was also stated that there was a leak at the luer connector.There was nothing unusual observed on the device prior to use.The catheter was not repaired.There was no instrument used to loosen or tighten the device.The cleaning agent used on the device was iodophor.Tego was not utilized.The insertion site was not treated prior to product placement.There was no other product being utilized with the device.There was no excessive force used on the device.The method that was utilized to tighten the adapters was by hand.The catheter was not replaced.Aside from the reported issue, there were no other visible defects/damages found on the product at the time of the event.There was no remedial action taken.The treatment was completed.There was no blood loss, and no blood transfusion was required due to the event.The catheter had been in place for 2 months.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted a crack on the venous luer adapter, with no other abnormalities observed on the device.It was reported that the luer adapter was leaking and cracked.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|