• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION, INC. AUTOPULSE® RESUSCITATION SYSTEM; CARDIAC RESUSCITATOR, LINE-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION, INC. AUTOPULSE® RESUSCITATION SYSTEM; CARDIAC RESUSCITATOR, LINE-POWERED Back to Search Results
Model Number 100
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported complaint of the autopulse platform (sn (b)(6) displayed the "system error, out of service, revert to manual cpr" error message was confirmed during initial functional testing and archive data review.The root cause of the system error was due to the failure of the processor board, likely attributed to the device's aging; the device's life expectancy is 5 years.The autopulse platform was manufactured in april 2013 and is over 12 years old.The archive data review showed the occurrence of the system error, user advisory (ua) 136 (internal parameter corrupted) error message.During the initial functional test, the platform displayed a "system error, out of service, revert to manual cpr" error message during power on, thus confirming the reported complaint.To remedy the error message, the failed processor board needs to be replaced.The last autopulse 1.1 was manufactured in 2009.As of july 2016, zoll announced the end of life for autopulse 1.1.Many important components used in autopulse 1.1 are no longer available, further restricting our ability to service.Zoll offered the customer a replacement device.Therefore, no service was performed, and the autopulse platform will be scrapped at zoll.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaints reported for the autopulse platform with serial number (b)(6).
 
Event or Problem Description
During the shift check, the autopulse platform (sn (b)(6) repeatedly displayed the "system error, out of service, revert to manual cpr" message.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION SYSTEM
Common Device Name
CARDIAC RESUSCITATOR, LINE-POWERED
Manufacturer (Section D)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave.
san jose, CA 95131
4084192347
MDR Report Key22503386
Report Number3010617000-2025-00471
Device Sequence Number18298863
Product Code DRM
UDI-Device Identifier00849111000277
UDI-Public00849111000277
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100
Device Catalogue Number8700-0700-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Initial Report FDA Received Date07/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-