| Model Number |
D141 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this device record a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data was sent for engineering analysis.Analysis showed that the low voltage fault is valid and confirmed that the power consumption is increasing in a gradual fashion.This device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.External visual inspection of the device revealed no anomalies.Review of the device memory confirmed that a low voltage alert was recorded.There were no other suspicious faults or resets.The battery voltage level upon receipt in the laboratory was sufficient to ensure therapy availability/delivery while the device was implanted.The device case was then opened to facilitate analysis of the internal components.The battery was separated from the other device electronics, and the overall current draw of the circuitry was measured.A normal current drain was observed within the circuitry.Detailed battery analysis determined that a latent current leakage path had occurred within the battery itself, resulting in a partial depletion of the battery.This behavior is caused by a latent electrical leakage path that develops over time between the anode and cathode within the device battery.In certain circumstances, lithium metal ions can re-plate to form clusters that may result in an internal current leakage.
|
| |
|
Event or Problem Description
|
|
It was reported that this device record a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data was sent for engineering analysis.Analysis showed that the low voltage fault is valid and confirmed that the power consumption is increasing in a gradual fashion.This device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|