| Model Number |
EGIA45AMT |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sigpshell, sig power sigpshell control shell (lot#:unknown), sigadaptstnd, sig power sigadaptstnd linear adapter (serial#(b)(6).Sigphandle, sig power sigphandle handle (serial#:unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter during an appendectomy, when attempting to fire, the shell and adapter detached.It was noted that the connection between the shell and adapter was not loose.Afterwards even after attaching the adapter, the firing did not return so it was released using the manual adapter.Another handle and adapter were used to resolve the issue.There was no patient injury.
|
| |
|
Event or Problem Description
|
|
According to the reporter during an appendectomy, when attempting to fire, the shell and adapter detached.It was noted that the connection between the shell and adapter was not loose.Afterwards even after attaching the adapter, the firing did not return so it was released using the manual adapter.Another handle, shell, adapter and reload were used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.H3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the firing did not return.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|