BARD ACCESS SYSTEMS, INC. POWERPICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
1174108D4 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Our vascular access team attempted to flush a "bard powerpicc" [peripherally inserted central catheter] that had been placed 24 hours earlier, with no luck.The rn was not able to flush the picc or have blood return once accessing the picc line.Patient had a right humerus x-ray, which showed the picc catheter double looped on itself at the patient's mid humerus.Patient underwent a picc underwire exchange in interventional radiology without issues, which fixed the double loop.
|
| |
|
Search Alerts/Recalls
|
|
|