| Catalog Number |
20123607 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
07/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer has indicated that the product will not be returned to zimmer biomet for investigation, as it was retained by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a left hip revision due to poly disassociation from the cup and the patient did not have a functioning poly for their hip.The liner was explanted.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: d9; g3; h2; h3; h4; h6; h10.Visual examination of the returned product identified the liner has scratches, gouges, and wear lines all around the outside surfaces.There are multiple portions of the outside rim lock and anti-rotation scallops worn completely off the liner.The top outside edge of the high wall appears heavily worn.There are multiple gouges on and near the high wall of the i.D.Surface.No other damage was noted during visual evaluation.Where measured, the device was conforming.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.Complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|