MAUDE Adverse Event Report: GYRUS ACMI, INC. 365 MICRON TFL SINGLE USE FIBER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. 365 MICRON TFL SINGLE USE FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number TFL-FBX365S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported during a bladder stone lithotripsy procedure utilizing a rigid laser cystoscope, the surgeon noted the absence of an aiming beam upon insertion of the laser fiber into the patient.The fiber was subsequently withdrawn for inspection, revealing a fractured tip and additional damage approximately one foot proximal to the distal end.The compromised fiber was replaced with a new one before proceeding with the procedure.There was no report of patient harm.
 
Event or Problem Description
No additional information received from the customer.
 
Additional Manufacturer Narrative
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: d8, d9, h2, h3, h6, h11 the device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause not established.Olympus will continue to monitor the performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
365 MICRON TFL SINGLE USE FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22506029
Report Number3011050570-2025-00725
Device Sequence Number12959020
Product Code GEX
UDI-Device Identifier00821925044005
UDI-Public00821925044005
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K183647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFL-FBX365S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/20/2025
Supplement Date Received by Manufacturer01/13/2026
Initial Report FDA Received Date07/15/2025
Supplement Report FDA Received Date02/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-