| Catalog Number |
QD8 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device had cosmetic damage, was loose, was difficult to assemble/disassemble, made unexpected noise, was vibrating and heated up.The device also failed pretests for visual assessment, nose tube assembly, thimble set screw, lock operation, vibration and temperature assessment.The assignable root cause was traced to component failure due to wear.
|
| |
|
Event or Problem Description
|
|
It was reported that the attachment device rotation was defective.It was unknown when the event occurred.There were no reports of delays to a surgical procedure.During in-house engineering evaluation, it was determined that the device had cosmetic damage, was loose, was difficult to assemble/disassemble, made unexpected noise, was vibrating and heated up.The device also failed pretests for visual assessment, nose tube assembly, thimble set screw, lock operation, vibration and temperature assessment.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2025.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|