ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC Back to Search Results
Catalog Number 705.44P
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that an ophthalmic forceps did not close properly during macular pucker peel surgery.Surgery was completed on the same day after replacing a product with new one.No patient impact was reported.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in sections d.9, h.3, h.6 and h.11.The returned instrument was received in its inner and outer blister and covered with the tyvek lid foil showing the lot information.The forceps was returned in an open state, and it shows clear signs of surgery residues.The returned instrument was visually inspected with the aid of a photomicroscope using various magnifications.The visual inspection revealed no immediate issues such as bumps or deformation of the forceps arms.The device was then functionally tested, it was noticed that no actuation of the forceps is possible.The actuation mechanism itself is still in working condition, but the forceps stuck in a close position.The forceps can only get opened again by manually pushing down the metal shaft, indicating that the bonding connection between the shaft and stiffening sleeve got broken.As the blister was opened by the customer, the product was handled.Therefore, the point in time when the malfunction occurred can no longer be determined, nor can the strain put on the device be reconstructed.The customer¿s complaint is confirmed, but a root cause for the connection failure cannot be determined.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Non-conformance report provided by originator.A non-conformance based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint it cannot be determined how or where the damage to the sample occurred.Therefore, the root cause for the customer's reported event could not be determined.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP FORCEPS
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22506642
Report Number3003398873-2025-00324
Device Sequence Number9403192
Product Code HNR
UDI-Device Identifier07612717070622
UDI-Public(01)07612717070622(17)270131(11)240206(10)16J1PV
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number705.44P
Device Lot Number16J1PV
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/10/2025
Supplement Date Received by Manufacturer08/07/2025
Initial Report FDA Received Date07/15/2025
Supplement Report FDA Received Date08/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/06/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-