• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4.5 HEALIX PEEK ANCH.W/OCORD; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4.5 HEALIX PEEK ANCH.W/OCORD; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 222205
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: e1: the reporter¿s complete facility address was not provided.The product has not returned to j&j medtech orthopaedics, however a photo was provided for review.Visual inspection of the returned photo found that the anchor was still attached to the inserter and broken at the distal end, the anchor had foreign matter, presumably biological matter.No other anomalies were noted.The overall complaint was confirmed as the observed condition of 4.5 healix peek anch.W/ocord would have contributed to the complained device issue.Based on the investigation findings, the potential cause for the broken anchor can be traced to off axis insertion and levering during insertion.As per ifu- 109139 inserting the awl or drill to less than the specified depth, axial misalignment or levering with the anchor upon insertion may result in anchor fracture.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
It was reported that during a rotator cuff repair procedure, the the 4.5 healix peek anch.W/ocord device anchor was broken off, removed all the broken parts.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 HEALIX PEEK ANCH.W/OCORD
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22507703
Report Number1221934-2025-02578
Device Sequence Number13077632
Product Code HWC
UDI-Device Identifier10886705007530
UDI-Public(01)10886705007530(10)250L697(17)261031
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K071481
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number222205
Device Lot Number250L697
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2025
Initial Report FDA Received Date07/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-