COVIDIEN LLC NELLCOR; OXIMETER Back to Search Results
Model Number PMC10GE360N
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device exhibited a faulty cord error, failed to detect a good plethysmographic waveform, and displayed random drops in spo2 readings below the patient¿s baseline.There was a discrepancy in oxygen saturation readings, with 96% recorded in triage and 83% on the monitor, and no good pleth waveform was detected with multiple pulse oximeter probes and sites.Audible and visual alarms for low oxygen saturation were activated.Other pulse oximeter probes were attempted on different fingers and the forehead without improvement.The cord was then swapped out, and accurate pleth and spo2 readings matching the triage value were confirmed.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d4 (udi), g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device exhibited a faulty cord error, failed to detect a good plethysmographic waveform, and displayed random drops in spo2 readings below the patient¿s baseline.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
COVIDIEN LLC
6135 gunbarrel ave
boulder,co
Manufacturer (Section G)
COVIDIEN LLC
6135 gunbarrel ave
boulder,co
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22509358
Report Number2936999-2025-00381
Device Sequence Number9303445
Product Code DQA
UDI-Device Identifier10884521816015
UDI-Public(01)10884521816015(11)231227(21)ST23520023(20)02
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K172482
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMC10GE360N
Device Catalogue NumberPMC10GE360N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Supplement Date Received by Manufacturer04/22/2026
Initial Report FDA Received Date07/15/2025
Supplement Report FDA Received Date05/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
Patient Weight86 KG
-
-