|
MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHX630 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient implanted with a 630cc mentor memorygel xtra breast implant experienced a rupture and an implant site hematoma on the right side post-operatively.The patient required surgical intervention to treat the seroma via aspiration.As a result, the patient underwent explantation on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted visual inspection, leak testing, and microscopic examination of the returned device.Visual inspection of the returned device determined that a tear was observed on the anterior view of the breast implant, measuring approximately 0.7 cm.Leak testing was performed, in accordance with mentor procedures, and no additional areas of gel exposure were detected during the analysis.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.Hematoma is a collection of blood within the space around the implant.Symptoms from hematoma may include swelling, pain, and bruising.If a hematoma occurs, it will most likely occur soon after surgery; however, it can occur at anytime, such as after an injury to the breast.Small hematomas may be absorbed by the body, while others may require surgery for drainage.Hematoma is a known complication associated with this procedure and is referenced in our product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|