H3: product analysis as found condition: two concerto devices were returned for analysis within a shipping box, within a sealed plastic biohazard pouch and within an opened concerto inner pouch.One device was returned further within a dispenser coil and partially within an introducer sheath with the implant coil and distal pusher already deployed out of the sheath.The second device was returned further within a syringe.The unknown non-medtronic micro catheter used in the event was not returned for analysis.Visual inspection/damage location details: (first coil) no damages or irregularities were found with the introducer sheath.The concerto pusher was found bent at ~49.4cm from the proximal end and found broken at ~141.7cm from the proximal end with the two segments retained by the inner wire.The concerto implant coil was found damaged outside its introducer sheath.(second coil) the pusher and introducer sheath were not returned.The already detached implant coil was found stretched and damaged with the polypropylene filament unbroken.Testing/analysis: (first coil) the concerto coil could not be retracted back within its introducer sheath as it was found to be stuck.The implant coil could not be used for resistance testing with an in-house micro catheter due to the damaged condition.(second coil) the concerto coil could not be used for resistance testing as the implant was already detached and due to the damaged condition.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil resistance/stuck at cath.Hub¿ was likely to have occurred as the damages found for both devices is consistent with resistance.Possible causes for coil resistance at catheter hub are, but not limited to, failure to hydrate the coil prior to use, failure to utilize continuous flush, failure to use a compatible micro catheter for delivery, damaged micro catheter, and use of an improper hubbing technique (over tightening of the rhv/sheath not firmly seated into hub).Customer reported devices were prepared per ifu, continuous flush was used, and no damage to coil/catheter.Therefore, the likely cause could not be determined.The first coil was found damaged with a bent/broken pusher, and the second coil was found damaged/stretched.It is possible the damage occurred during the attempt to push the coil into the micro catheter hub against the reported resistance.Other possible causes of coil stretching are lack of hydration before procedure, user does not maintain continuous flush, coil not retracted in a one-to-one motion with the implant pusher during repositioning or pushwire rotation.The second implant was found already detached.Customer did not report detaching the device; therefore, it is likely the first coil belongs to pli-10.As the unknown non-medtronic micro catheter used during the event was not returned for analysis, any contribution of the micro catheter towards the resistance could not be determined.Customer reported successfully deploying a different coil; therefore, it is likely in good working condition.As the model and lot numbers were not reported, compatibility could not be assessed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|