OLYMPUS WINTER & IBE GMBH BIOPSY FORCEPS, SPOON TYPE, OPTICAL, FOR 12° TELESCOPE; A20712A
|
Back to Search Results |
|
| Model Number |
A20712A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the subject device exhibited a broken wire in the jaw.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely that the wire broken in the jaw as a result of stress.However, a definitive root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|