• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX MEDICAL; IMAGINA COLONOSCOPE (US ONLY)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX MEDICAL; IMAGINA COLONOSCOPE (US ONLY) Back to Search Results
Model Number EC38-I10CL
Medical Device Problem Code Optical Obstruction (3002)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb shadow in image.Based on the result, we concluded that it was caused due to the excessive force applied on the cmos module with drive pcb.In addition, our technician confirmed that the insertion flexible tube coating damage, and the air tube leak; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure(shadow in image).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX MEDICAL
Common Device Name
IMAGINA COLONOSCOPE (US ONLY)
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key22516770
Report Number9610877-2025-52542
Device Sequence Number8058807
Product Code FDF
UDI-Device Identifier04961333248513
UDI-Public(01)04961333248513(11)220518
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K190805
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEC38-I10CL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2025
Initial Report FDA Received Date07/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-