| Model Number |
8229707 |
| Medical Device Problem Codes |
False Positive Result (1227); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting guidance from fda's 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during thyroidectomy procedure the site experienced excessive noise from channel 1.This was not experienced until about 3 hours into the case.Upon running an electrode check, channel 2 was intermittently failing and passing.Upon removing the tube, user noted the tube to be slightly bent.The procedure was completed by reported devices.There was no patient impact.
|
| |
|
Search Alerts/Recalls
|