• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC DOMINICANA NIM® EMG - ENDOTRACHEAL TUBE - TRIVANTAGE®; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC DOMINICANA NIM® EMG - ENDOTRACHEAL TUBE - TRIVANTAGE®; STIMULATOR, NERVE Back to Search Results
Model Number 8229708
Medical Device Problem Codes Image Display Error/Artifact (1304); Inappropriate Audible Prompt/Feedback (2280)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting guidance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that site is using a nim that will not pass its electrode check.Upon plugging in the patient simulator, the electrode check passed normally.Adjusting the tube allowed the electrode check would pass intermittently.The site already attempted to replace both the tube and the ground electrode.They changed from a 7.0 mm tube to an 8.0 mm tube.The electrode check does not pass but it will stimulate.Even threshold is on but not causing the issues.They used the mini screen attached to a metal pole.When they touch the patient interface it made very excessive noise as if alot of metal touched it ( its a mix between static and like metal clanging).There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® EMG - ENDOTRACHEAL TUBE - TRIVANTAGE®
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR  11500
Manufacturer (Section G)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR   11500
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key22517859
Report Number9612501-2025-01868
Device Sequence Number15611563
Product Code ETN
UDI-Device Identifier00763000033965
UDI-Public(01)00763000033965(17)261124(10)0225551378
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K112686
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Remedial Action Notification,Recall
Type of Report Initial
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8229708
Device Catalogue Number8229708
Device Lot Number0225551378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2023
Initial Report FDA Received Date07/16/2025
Date Device Manufactured11/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0696-2024
Patient Sequence Number1
Patient SexUnknown
-
-