| Model Number |
55750026545 |
| Medical Device Problem Codes |
Backflow (1064); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E: first name and last name of initial reporter is unknown.G2: country of origin is japan.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via manufacturer representative regarding product used in spinal therapy for compression fracture.It was reported that to prevent cement leakage into the screw head, the inline method was used to check for any axial misalignment, and the delivery guide was inserted.The passage was confirmed using a 1.5k wire, and under the c-arm, the flow of cement into the vertebral body was verified.After cement filling, it was confirmed that the tip did not come off when removing the delivery guide.Attempts were made to remove the cement using a chisel and detach the tip with a punch, but these efforts were abandoned.Since attention was paid to both the viscosity and the filling speed of the cement, the cause of this incident could not be determined.There was a delay of less than 60 minutes and no further complications or symptoms reported.Additional information was received via manufacturer representative that there is no extravasation of cement is observed.There is no allegation of cement but this event is reported for screw malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received via manufacturer representative that there are no complaints on fns tip and cause of the event is unknown if tip might not be properly dropped into the screw head.There are complaints that the delivery guide is difficult to use.
|
| |
|
Event or Problem Description
|
|
Additional information was received via manufacturer representative that there was cement leaked from screw head and backflow observed in fns screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Radiographic image: antero posterior x-ray l4-s1 interbody fusion, cement augmentation of pedicle screws was unable to visualize as described broken hardware.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|