It was reported to intervascular that, on (b)(6) 2024 a patient was treated in the operating room following an ascending aortic aneurysm.During this procedure, the following devices were used: bioglue 10ml syringe (reference: (b)(4); supplier: (b)(4), cardioroot aortic vascular prosthesis with bulb d32mm l15cm (hewroot0032) (reference: (b)(4); supplier: (b)(4), avalus bovine aortic valve d27mm (reference: (b)(4); supplier: (b)(4).Since this procedure, the patient has suffered from chronic hypereosinophilia.An etiological assessment was performed by the hospital and no cause was found.The parasitological examination was negative and the drug-related cause ruled out with pharmacovigilance.
|
|
(4111) attempts to contact the initial reporter in order to get more information about the patient information the involved products were requested and responses about the multiple procedures performed were provided as follows: on (b)(6) 2025, the patient was hospitalized for what is described as a general deterioration with suspicion of mucormycosis.Aspergillus, mycoplasma and chlamydia were ruled out.An echocardiogram performed identified prosthetic valve thickening and suspected endocarditis, but blood cultures were negative.Ceftriaxone and levofloxacin were administered but discontinued due to skin rash.The patient was discharged on (b)(6) 2025, still weakened, dyspneic, with fever and tachycardia, having lost 6 kg.The patient was again hospitalized (b)(6) 2025, with continued symptoms, dysuria, anorexia, sleep disturbances, bilateral rales and chronic cough.A new ct scan showed multiple alveolar condensations, bilateral pleural effusion, no prosthetic infection, but persistent mediastinal-hilar uptake.A slight improvement of the basal alveolar consolidations was observed.Bilateral pneumonia with actinomyces was diagnosed.Cardiovascular prostheses were studied (echocardiogram, blood cultures, transesophageal pacing) to rule out infection, no evidence was found to sustain this hypothesis.The cause of chronic hyper eosinophilia was not determined, continued monitoring by infectious diseases.Amoxicillin i.V.Was administered, isavuconazole was discontinued.The patient¿s lower left molars were surgically removed.Severe malnutrition treated with nasogastric tube feeds.Patient was discharged (b)(6) 2025.During an office visit (date unknown) it was reported that enteral nutrition was discontinued (b)(6) 2025.The patient presented without fever or chills, continues taking amoxicillin p.O.A ct scan showed a complete regression of the pleural effusion, and slight regression of the mediastinal lymphadenopathies, as well as an improvement of the alveolar consolidations.The patient¿s current status is unknown.(4112/213) the case and its investigation have been reviewed by the medical affairs department whose conclusion is as follows: hypereosinophilia can be secondary to sensitization caused by an allergic reaction, autoimmune disorders, parasitic infections, or drug hypersensitivity.The implantation of the intergard cardioroot aortic vascular prosthesis, the avalus bovine aortic valve, or bioglue cannot be ruled out as the cause of hypereosiniphilia." (4110/213) the actual occurrence rate was calculated for allergic reaction (including this complaint) on the same product family (intergard /hemagard standard grafts and patches) from (b)(6) 2024 to (b)(6) 2025.The calculated occurrence rate is (b)(4) for the identified risk, which is below to the anticipated occurrence rate (upper control limit rate) of (b)(4).The risk assessment is still adequate.(4315) the cause of the event remains unknown.However, the conducted investigation suggests that the product was not defective at the time of manufacturing.The medical review concluded that chronic hypereosinophilia can be secondary to sensitization caused by an allergic reaction, autoimmune disorders, parasitic infections, or drug hypersensitivity.The implantation of the intergard cardioroot aortic vascular prosthesis, the avalus bovine aortic valve, or bioglue cannot be ruled out as the cause of hypereosiniphilia.
|