| Catalog Number |
365967 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported prior to use of bd microtainer® sst¿, 1 tube had the gel at the top of the tube instead of the bottom.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples or photos for investigation.Therefore, 50 retention samples were evaluated by visual examination and no issues were observed relating to separator position as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode additive abnormality.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported prior to use of bd microtainer® sst¿, 1 tube had the gel at the top of the tube instead of the bottom.There was no health impact or consequences reported.
|
| |
|