OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O. HF-RESECTION ELECTRODE "PLASMABAND", BAND, MEDIUM, 24 FR., 12°-30°, ESG TCRIS; WA47723S Back to Search Results
Model Number WA47723S
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the resection electrode plasmaband had the hook bent in an unknown and abnormal way.The issue was found during a myomectomie therapeutic procedure.The intended procedure was completed with an olympus back-up device.There were no reports of patient injury or harm.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
No additional information received from the customer.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation: the most probable cause of the complaint was traced to component failure, which is expected or random component failure without any design or manufacturing issue.Deformation problem identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMABAND", BAND, MEDIUM, 24 FR., 12°-30°, ESG TCRIS
Common Device Name
WA47723S
Manufacturer (Section D)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ  75124
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ   75124
Manufacturer Contact
todd brill
801 west park drive
westborough, MA 01581
5082077661
MDR Report Key22520024
Report Number3007960282-2025-00104
Device Sequence Number12919890
Product Code FAS
UDI-Device Identifier14042761085561
UDI-Public14042761085561
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171965
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA47723S
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/25/2025
Supplement Date Received by Manufacturer01/05/2026
Initial Report FDA Received Date07/16/2025
Supplement Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-