• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Catalog Number 130790030
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that patient felt a pop in their shoulder and went in for a checkup.X-rays showed the glenosphere had broken loose, leaving a couple broken screws in the glenoid.When revising, surgeon felt the stem was stable, so surgeon didn¿t want to revise the humeral component.Surgeon was able to remove the glenosphere and broken screws and successfully prep the glenoid for a new baseplate and glenosphere.A new epi and poly were implanted as well.Stability of the shoulder was confirmed and the closing process began.There was no delays or adverse events took place.Doi: (b)(6) 2023, dor: (b)(6) 2025, affected side: right shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: d10 (concomitant).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X30MM
Common Device Name
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22520571
Report Number1818910-2025-12234
Device Sequence Number20847744
Product Code HSD
UDI-Device Identifier10603295027843
UDI-Public(01)10603295027843(17)20270731(10)5419459
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number130790030
Device Lot Number5419459
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer07/21/2025
Initial Report FDA Received Date07/16/2025
Supplement Report FDA Received Date07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND METAGLENE.; DXTEND MOD CENT EPI 2 HA.; DXTEND SCREW LOCK D4.5X30MM.; DXTEND SCREW NO LOCK D4.5X18MM.; DXTEND SCREW NO LOCK D4.5X18MM.; DXTEND STAND PE CUP D42 +3MM.; XTND GLENO D42MM +2MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-