| Model Number |
103-0606-200 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Ischemia Stroke (4418); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2025
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Information received indicated pli30 and pli40 are no longer reportable.Since a regulatory report has been sent gch functionality does not allow the complaint flag to be changed to no.Should further information make the event reportable again the event will be reopened and a regulatory report will be sent.See section h.6.For corrected codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id sfr4-6-40-10 (b791192); implant date n/a; explant date n/a product id 105-5081-153 (b813957); implant date n/a; explant date n/a product id sfr4-4-40-10 (b777662); implant date n/a; explant date n/a medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a solitaire x stent retriever (sfr4-6-40-10, lot: b791192) that broke during a procedure.It was reported that patient was undergoing a mechanical thrombectomy procedure to treat a non-hemorrhagic stroke.The solitaire and all accessory devices were prepared as indicated in the instructions for use (ifu).The solitaire was deployed for clot retrieval.However, during retrieval, there was resistance in the distal end of the catheter (rebar, 105-5081-153, lot: b813957) and the solitaire broke and remained in the clot.The surgeon used another solitaire x (sfr4-4-40-10, lot: b777662) to retrieve the broken solitaire along with the clot.It appears the broken device was successfully retrieved.However, it was noted that the patient ultimately had an outcome of death.
|
| |
|
Search Alerts/Recalls
|