• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 87435
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during pre-test, the endotracheal tube's cuff was herniated, as shown in the image attached.There was no patient involved.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the cuff presented a distortion.An inflation/deflation test was performed air was applied to the cuff using a syringe but no leaks were observed.It was reported that the 7.5mm endotracheal tube's cuff pressure was noted to be deflated.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not overinflate the cuff.It is recommended that the cuff pressure should not exceed 25cmh2o.Overinflation can result in tracheal damage, rupture of the cuff with subsequent deflation, or in cuff distortion which may lead to airway blockage.Test the cuff, pilot balloon and valve for integrity by inflation prior to use.Insert a luer tip syringe into the cuff inflation valve housing and inject enough air to fully inflate cuff.After test inflation, completely evacuate the air.Various bony anatomical structures (e.G., teeth, turbinates) within the airway or any intubation tools with sharp surfaces might jeopardize cuff integrity.Damage to the thin-walled cuff during insertion will subject the patient to the risk of extubation and reintubation.If the cuff is damaged, the tube should not be used.Deflate the cuff prior to repositioning the tube.Movement of the tube with the cuff inflated could result in patient injury, requiring possible medical intervention or damage to the cuff.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22521812
Report Number9681384-2025-00755
Device Sequence Number8078946
Product Code BTR
UDI-Device Identifier10884521816282
UDI-Public10884521816282
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number87435
Device Catalogue Number87435
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer12/10/2025
Initial Report FDA Received Date07/16/2025
Supplement Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-