This supplemental report is being submitted to provide the legal manufacturer¿s final investigation.Additional data: d8, h3, h4.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, the issue is attributed to electrical problems, however, a definitive root cause could not be established.It is likely the following led to the malfunction: damage or deterioration of parts, environment problem like as dust/dirt/humidity, optical problem, overheating problem, and electrical problems like as signal loss or open circuit.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|