• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP TM 6.0MM MTX FULL, 10; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP TM 6.0MM MTX FULL, 10; DENTAL IMPLANT Back to Search Results
Catalog Number TMT6B10
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
Date of Event 06/03/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth site #14 was removed due to peri-implantitis.Symptoms as a result of the event: pain.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided.G4: additional pma/510(k) number ¿ k113753|k112160.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP TM 6.0MM MTX FULL, 10
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22522739
Report Number0002023141-2025-01967
Device Sequence Number8058037
Product Code DZE
UDI-Device Identifier00889024018969
UDI-Public(01)00889024018969(17)250721(10)1238356
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K132258
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/21/2025
Device Catalogue NumberTMT6B10
Device Lot Number1238356
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-