As reported, during a procedure involving insertion of a tunneled catheter on the left side, a bentson straight fixed core wire guide "sheared off".The wire was not altered prior to use.Per the reporter, there was no issue inserting the wire; however, the user felt "some" resistance.When the physician pulled the wire out of the catheter, the outside of the wire "sheard off".A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures or experience any adverse effects due to this event.
|
|
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during a procedure involving insertion of a tunneled catheter on the left side, a bentson straight fixed core wire guide "sheared off".The wire was not altered prior to use.Per the reporter, there was no issue inserting the wire; however, the user felt "some" resistance.When the physician pulled the wire out of the catheter, the outside of the wire "sheard off".A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures or experience any adverse effects due to this event.Corrected information: d4 = udi.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.A review of the device history record and complaint history found no discrepancies or additional relevant complaints on the final or sub-assembly lots.The customer followed relevant instructions in the ifu (instructions for use).A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr and dhr suggests that there is evidence the device was manufactured to specification.Based on the information provided and the results of the investigation, cook has concluded that a component failure, unrelated to manufacturing or design deficiencies, contributed to this event.The risk analysis for this failure mode was reviewed, and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|