| Model Number |
15100 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The screw has not returned for evaluation.Radiographic images were provided which confirm the event of two set screw backouts and a rod slippage.A review of the device history records could not be performed as the identifying lot number was not provided.Alphatec spine is submitting this report to comply with the fda regulations 21 cfr 803.Alphatec spine has made reasonable efforts to provide as much relevant information as available.Any field that is left blank was not known at the time of this submission.If additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
A patient underwent a spinal surgery on an unknown date.About 6 weeks postoperatively radiographic images revealed that two set screws had become disengaged, and the rod slipped.The patient is asymptomatic, and there are no plans for revision surgery.This is report 1 of 2.
|
| |
|