CONMED UTICA VCARE; CANNULA, MANIPULATOR/INJECTOR, UTERINE Back to Search Results
Catalog Number 60-6085-202A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reported event of ¿balloon broke off inside the abdomen and the blue and green parts snapped off inside the vagina¿ is confirmed.The device will not be returned for evaluation; however, photographic evidence has been provided.Root cause cannot be determined, however, based ifu; a possible cause of this event could be excessive force.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of 1 device for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 19 complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised to re-attach the syringe to the luer connector at the end of the pilot balloon; fully aspirate the air from the intrauterine balloon to deflate.This will allow the intrauterine balloon to be removed from the uterus.Unlock the locking mechanism by turning the thumbscrew counter-clockwise (anti-clockwise) and retract to the handle.A.Swipe a finger around the edge of the vaginal cup to separate the tissue from the cup to prevent tissue damage b.Fully retract the vaginal cup to the handle.Carefully remove the device from the vagina.Do not use excessive force to avoid traumatizing the vaginal canal.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Event or Problem Description
The customer reported that the that the 60-6085-202a, lg, uterine manipulator device was being used during an unknown procedure on (b)(6) 2025 when it was reported ¿v care that was used today that broke off inside the patient once the uterus had been disconnect from all tissue.The balloon broke off inside the abdomen and the blue and green parts snapped off inside the vagina.Luckily this was noticed, and all broken parts were retrieved.¿.There was no report of injury, medical intervention, or extended hospitalization for the patient or user.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCARE
Common Device Name
CANNULA, MANIPULATOR/INJECTOR, UTERINE
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key22529340
Report Number1320894-2025-00176
Device Sequence Number15429676
Product Code LKF
UDI-Device Identifier10653405003809
UDI-Public(01)10653405003809(17)270120(10)202501201
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K142716
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number60-6085-202A
Device Lot Number202501201
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Initial Report FDA Received Date07/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-