| Catalog Number |
250610005 |
| Medical Device Problem Codes |
Break (1069); Nonstandard Device (1420); Defective Device (2588); Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported fragments/burrs were discovered inside holes of the instruments during inspection.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary.According to the information received, "instruments enter barts health qa process before acceptance and the following issues" the product was not returned to j&j medtech orthopaedics, however photos were provided for review.See attachment (b)(4) - user report on 2025_006_030_501_004 _mhra ref_ 36036076-aicxml_ (3).Msg.The photo investigation revealed signs of burr on the insertion holes of the attune rev fb trial size 5.However no signs of breakage were not found.The observed condition may be due to multiple attempts at assembly and disassembly of the low-profile tibial pin, however a definite root cause cannot be established with the photos provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.A manufacturing record evaluation was performed for the finished device product description: attune rev fb trial size 5 product code: 250610005 lot number: so2066586 and no non-conformances or manufacturing irregularities were identified.The overall complaint was confirmed as the observed condition of the attune rev fb trial size 5 would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot.1) quantity manufactured: 108.2) date of manufacture: 10/3/2024.3) any anomalies or deviations identified in dhr: none.4) expiry date: n/a.5) ifu reference: ifu ns reuse inst rev c.Device history review.1) quantity manufactured: 108.2) date of manufacture: 10/3/2024.3) any anomalies or deviations identified in dhr: none.4) expiry date: n/a.5) ifu reference: ifu ns reuse inst rev c.H11 additional narrative: added: h6 (health effect - impact code).Corrected: h4, h6 (medical device problem code).
|
| |
|
Search Alerts/Recalls
|