| Model Number |
PCF-H180AI |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure found during investigation was confirmed.According to the investigation, foreign material was found in the scope air/water cylinder (tube)/tube.However, component analysis of the foreign material could not be identified.The root cause could not be identified.Based on the results of the investigation, it could not conclusively specify the root cause of the foreign material remained in the device.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the colonovideoscope exhibited foreign material in the scope air/water cylinder (tube)/tube.There was no patient involvement.
|
| |
|