• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 36IN 4-0 D/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 36IN 4-0 D/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 8521H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an aortic root enlargement procedure on (b)(6) 2025 and suture was used.During the procedure, suture breakage during knotting.It was reported that total 6 products occurred suture breakage during knotting issue.Changed another lot number product to complete the aortic root enlargement surgery.There is no report on patient's injury.No additional information could be provided.There will return 25 representative samples for analysis.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H3 investigation summary: the product was returned to ethicon for evaluation.One open box with twenty-five unopened samples that pertains to product code 8521h.As per the sampling plan, a visual inspection was performed on thirteen samples, and no defects were found on the packages.The packets were opened, and the swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along of the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.Also, a functional test was performed using instron equipment and tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 36IN 4-0 D/A SH
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
*   00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22530364
Report Number2210968-2025-08245
Device Sequence Number7932124
Product Code GAW
UDI-Device Identifier10705031016538
UDI-Public(01)10705031016538(17)290731(10)102HCQ
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8521H
Device Lot Number102HCQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Initial Report FDA Received Date07/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-