Zimvie did not receive one (1) bopt4310, (3i t3® tapered implant 4/3 x 10mm) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported events could not be performed due to the nature of the device & events.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 2120015980.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120015980 for similar events and one (1) other complaint was identified (b)(4).The customer did not submit images for the reported events.Based on the investigation and risk management file review as per rmf rm-00053-haz rev.12, the most likely root cause determined from the investigation is inadequate treatment planning.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported events are non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report g3: date received by manufacturer g6: checked "follow-up" h6: entered evaluation codes h10: added manufacturer narrative.
|