Model Number 3000TFX
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 07/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported and learned through medical records that a patient with a 21mm 3000tfx aortic valve is under consideration for a valve-in-valve procedure after an implant duration of 9 years, 6 months due to severe calcific stenosis.Patient presented with heart failure exacerbation.This is an 88-year-old female patient.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections: , g3, g6, h2, h6 (type of investigation) corrected sections: h6 (component code, investigation findings and investigation conclusions).Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration there is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years and hyperlipidemia, diabetes mellitus.A device history record (dhr) review is not performed because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
The procedure was completed.A supplemental report will be submitted if any additional information is provided.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported and learned through medical records that a patient with a 21mm 3000tfx aortic valve is underwent a valve-in-valve procedure after an implant duration of 9 years, 6 months due to severe calcific stenosis.Patient presented with heart failure exacerbation.The procedure is performed with 23mm 9750tfx transcatheter valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key22532691
Report Number2015691-2025-05826
Device Sequence Number18010716
Product Code DYE
UDI-Device Identifier00690103163643
UDI-Public(01)00690103163643(17)190510
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/10/2019
Device Model Number3000TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer08/19/2025
09/04/2025
Initial Report FDA Received Date07/17/2025
Supplement Report FDA Received Date08/21/2025
09/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/11/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexFemale
-
-