| Model Number |
90010200499991010 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing, which included all stress testing and troubleshoot testing without duplicating the customer's report.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Search Alerts/Recalls
|