|
C.R. BARD, INC. (BASD) -3006260740 POWERPICC HF (HIGH FLOW), 3CG; PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)
|
Back to Search Results |
|
| Catalog Number |
1385118Q |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by customer that the "buckling" was occurring quite frequently, and they had to adapt their dilation technique to accommodate for it (larger nick in the skin, stretch and pause before continuing, and holding much closer to hub of dilator).Most often, the dilator would just buckle and not create the rough edge, but sometimes it created the rough edge making it unusable.The dilators are unusable and had to be replaced mid-procedure no further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|