• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA Back to Search Results
Model Number OJR418VT
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr418vt optiflow junior 2 nasal cannula was detached from the swivel grip joint during use on a patient.There was no reported patient consequence.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject devices, ojr418vt optiflow junior 2 nasal cannulas were returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the evaluation of the returned devices, information provided by the healthcare facility and our knowledge of the product.Results: visual inspection of the two returned devices was performed.One cannula had one of its tubings detached from the cannula tube joint.The second device had one of its tubings detached from the connector end (swivel grip).Conclusion: based on the visual inspection of the two returned devices, information provided and our knowledge of the product, it is most likely that the reported event resulted from the subject device being subjected to excessive force.As part of the manufacturing process, all optiflow junior nasal cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken throughout each batch and functionally tested for retention strength between the assembled parts.If there are any failures the entire batch is put aside for further investigation.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: - appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death).- do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).- do not stretch or crush tube.- ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
 
Event or Problem Description
A healthcare facility in florida reported via a fisher & paykel healthcare (f&p) field representative that two ojr418vt optiflow junior 2 nasal cannulas were found broken during use on a patient.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER AND PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key22539073
Report Number9611451-2025-00649
Device Sequence Number12630867
Product Code BTT
UDI-Device Identifier09420012448811
UDI-Public(01)09420012448811(10)2103463311(11)241108(17)291108
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K222197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOJR418VT
Device Catalogue NumberOJR418VT
Device Lot Number2103463311
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2025
Supplement Date Received by Manufacturer10/01/2025
Initial Report FDA Received Date07/18/2025
Supplement Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RT232 MICROCELL CIRCUIT AND MR290 WATER CHAMBER KIT
Patient SexUnknown
-
-