• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC KYPHON® XPEDE¿ BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC KYPHON® XPEDE¿ BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number CX01B
Medical Device Problem Codes Coagulation in Device or Device Ingredient (1096); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6 - neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a kyphon and hvr bone cement with mixer having spinal therapy.It was reported that procedure started out normal with access via pedicles into vertebrae and then osteocool was performed on levels t5 and t11.Once the osteocool was done, the kyphoplasty with balloons was performed without any issues.There was difficulty mixing the bone cement due to lack of familiarity with the equipment.When administering the xpede into the bfds, the cement wasn't coming out and within about 1 minute of it being mixed it hardened in the bone filler and the mixer with plunger.A second xpede kit was opened when this happened and ended with the same result, which at that point a third xpede kit was opened with mixer and within a minute of mixing it, the cement had hardened and the scrub tech broke the top of the mixer off.Additionally, the hvr bone cement was unable to be delivered because it was too warm to work with upon arrival in the ir room.Multiple cement mixer kits (four) were rendered unusable during the procedure.The case was completed using an alternative product from another manufacturer.Hvr- cement hardened faster.Event occurred at back table when setting up the cement.There was no patient involvement reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON® XPEDE¿ BONE CEMENT
Common Device Name
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22540147
Report Number1030489-2025-02465
Device Sequence Number18125791
Product Code NDN
UDI-Device Identifier00643169097797
UDI-Public(01)00643169097797(17)280131(10)0012839147
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K102397
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCX01B
Device Catalogue NumberCX01B
Device Lot Number0012839147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2025
Supplement Date Received by Manufacturer09/19/2025
Initial Report FDA Received Date07/18/2025
Supplement Report FDA Received Date09/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-