| Model Number |
C01B |
| Medical Device Problem Codes |
Coagulation in Device or Device Ingredient (1096); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6 - neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a kyphon and hvr bone cement with mixer having spinal therapy.It was reported that procedure started out normal with access via pedicles into vertebrae and then osteocool was performed on levels t5 and t11.Once the osteocool was done, the kyphoplasty with balloons was performed without any issues.There was difficulty mixing the bone cement due to lack of familiarity with the equipment.When administering the xpede into the bfds, the cement wasn't coming out and within about 1 minute of it being mixed it hardened in the bone filler and the mixer with plunger.A second xpede kit was opened when this happened and ended with the same result, which at that point a third xpede kit was opened with mixer and within a minute of mixing it, the cement had hardened and the scrub tech broke the top of the mixer off.Additionally, the hvr bone cement was unable to be delivered because it was too warm to work with upon arrival in the ir room.Multiple cement mixer kits (four) were rendered unusable during the procedure.The case was completed using an alternative product from another manufacturer.Hvr- cement hardened faster.Event occurred at back table when setting up the cement.There was no patient involvement reported.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|