| Catalog Number |
08056960190 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 2 patients' samples tested with bil-d gen.2 (bild2) assay and 3 patients' samples tested with bilirubin total gen.3 (bilt3) assay and bil-d gen.2 (bild2) assay on a cobas pure c 303 analytical unit, a c 501 module and, an integra analyzer compared to the siemens method.This medwatch will apply to the bilt3 assay.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the bild2 assay.Patient 1: bild2: initial result: 0.894 mg/dl (tested on pure c 303).1st repeat result: 0.215 mg/dl (tested on cobas integra).2nd repeat result: 0.334 mg/dl (tested on cobas c501).3rd repeat result: 0.21 mg/dl (tested on siemens).Patient 2: bild2: initial result: 0.615 mg/dl (tested on pure c 303).1st repeat result: 0.342 mg/dl (tested on cobas integra).2nd repeat result: 0.468 mg/dl (tested on cobas c501).3rd repeat result: 0.24 mg/dl (tested on siemens).Patient 3: bilt3: initial result: 0.118 mg/dl (tested on pure c 303).1st repeat result: 0.113 mg/dl (tested on cobas integra).2nd repeat result: 0.135 mg/dl (tested on cobas c501).3rd repeat result: 0.21 mg/dl (tested on siemens).Bild2: initial result: 0.094 mg/dl (tested on pure c 303).1st repeat result: 0.122 mg/dl (tested on cobas integra).2nd repeat result: 0.119 mg/dl (tested on cobas c501).3rd repeat result: 0.10 mg/dl (tested on siemens).Patient 4: bilt3: initial result: 0.222 mg/dl (tested on pure c 303).1st repeat result: 0.181 mg/dl (tested on cobas integra).2nd repeat result: 0.205 mg/dl (tested on cobas c501).3rd repeat result: 0.27 mg/dl (tested on siemens).Bild2: initial result: 0.117 mg/dl (tested on pure c 303).1st repeat result: 0.120 mg/dl (tested on cobas integra).2nd repeat result: 0.134 mg/dl (tested on cobas c501).3rd repeat result: 0.07 mg/dl (tested on siemens).Patient 5: bilt3: initial result: 0.155 mg/dl (tested on pure c 303).1st repeat result: 0.162 mg/dl (tested on cobas integra).2nd repeat result: 0.160 mg/dl (tested on cobas c501).3rd repeat result: 0.22 mg/dl (tested on siemens).Bild2: initial result: 0.097 mg/dl (tested on pure c 303).1st repeat result: 0.101 mg/dl (tested on cobas integra).2nd repeat result: 0.116 mg/dl (tested on cobas c501).3rd repeat result: 0.05 mg/dl (tested on siemens).The customer questioned the initial bild2 results as they were higher than expected and did not match the patients' clinical status.No questionable results were reported outside the laboratory, and the samples were then repeated.The 3rd repeat results were deemed to be correct and reported to the medical personnel.
|
| |
|
Additional Manufacturer Narrative
|
|
The bild2 reagent lot number was 839949 with an expiration date of 28-feb-2026.The versions of bild2 and bilt3, including the lot numbers and expiration dates that were run on the cobas integra and the c501 analyzer, were not provided.The cobas pure c 303 analytical unit serial number was (b)(6).The serial numbers of the cobas integra and cobas c 501 analyzers were not provided.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
No hardware issue was present.The customer has not provided any additional information for the investigation.Different manufacturers use different standardization protocols for their assays; therefore, bilirubin results of different manufacturers are not directly comparable.Differences in the assay composition and method might lead to slightly different reactivity.The investigation did not identify a product problem.The specific cause of the event could not be determined.
|
| |
|
Search Alerts/Recalls
|