• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL, INC. PORTEX ABS PRO-VENT ARTERIAL BLOOD SAMPLING; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL, INC. PORTEX ABS PRO-VENT ARTERIAL BLOOD SAMPLING; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number 4698P-2
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 phone (b)(6).Customer provided lot number 6072428; however, the lot was unable to be verified.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event or Problem Description
It was reported that the vented cap was leaking, not sealing when in contact with blood which was confirmed that the users are twisting caps as they push down firmly to close, per guidance provided.There was no patient harm/adverse event.
 
Additional Manufacturer Narrative
No product was returned; however, photos of the device was provided for analysis.Functional and visual tests were performed, but the reported issue could be duplicated.The cause of the issue couldn't be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ABS PRO-VENT ARTERIAL BLOOD SAMPLING
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
ICU MEDICAL, INC.
10 bowman dr
keene NH 03431
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key22541423
Report Number3012307300-2025-08516
Device Sequence Number7029126
Product Code JKA
UDI-Device Identifier20351688032802
UDI-Public(01)20351688032802
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K952516
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4698P-2
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2025
Supplement Date Received by Manufacturer02/03/2026
Initial Report FDA Received Date07/18/2025
Supplement Report FDA Received Date02/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-