• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 50MM RT NARROW MAND; TOTAL TEMPOROMANDIBULAR JOINT PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 50MM RT NARROW MAND; TOTAL TEMPOROMANDIBULAR JOINT PROSTHESIS Back to Search Results
Catalog Number 01-6550
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/14/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an initial tmj procedure approximately 3 years ago.Subsequently, the patient has experienced infections over the past two years, claiming that the implants did not fit properly and have punctured their ear canal, contributing to infection.The patient was revised on an unknown date.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).D10: 50mm lft narrow mand cat# 01-6551 lot# 069750a.Tmj sm rt fossa comp cat# 24-6562 lot# 085160c.Tmj sm lft fossa comp cat# 24-6563 lot# 073870d.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a4, b3, b4, b5, b7, d4, d6b, e1, g3, g6, h2, h11.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial bilateral tmj procedure approximately 3 years ago.Subsequently, the patient alleges the left joint became infected shortly after implantation causing a hole in the ear canal.Due to this, the patient alleges a staph infection occurred.Therefore, the patient underwent a revision where all components were removed and replaced with custom devices.The patient alleges the revision surgeon noted when the original implants were initially implanted, that they did not fit properly and the initial surgeon used cement to force the implants to fit into the joint space.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event or Problem Description
Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.The reported event could not be confirmed due to lack of product/information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.The device history record was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported issue is attributed to a usage error.The ifu for these devices states bone cement or other grouting agents should not be used when implanting these devices as safety and efficacy have not been established for bone cement.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
50MM RT NARROW MAND
Common Device Name
TOTAL TEMPOROMANDIBULAR JOINT PROSTHESIS
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22541517
Report Number0001032347-2025-00278
Device Sequence Number7013946
Product Code LZD
UDI-Device Identifier00841036011895
UDI-Public(01)00841036011895(17)270107(10)077940A
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P020016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number01-6550
Device Lot Number077940A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2025
Supplement Date Received by Manufacturer07/24/2025
12/22/2025
Initial Report FDA Received Date07/18/2025
Supplement Report FDA Received Date08/08/2025
12/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age33 YR
Patient SexFemale
Patient Weight52 KG
-
-