Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening/tensioning.The dfu for this device, dfu-0147-eo revision 4, gives the following precaution: acl/pcl/btb/rt/abs/cl/cls/pf/pfc tightrope only: excessive force on the shortening suture strands and/or tensioning the shortening suture strands not in-line with the bone socket may break the strands and impair ability to fully seat the implant.No additional force on the shortening suture strands is required when the graft/wedge construct reaches the desired position in the femoral socket, and the graft stability is verified by pulling distally on the graft.
|