• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. LO-PRO SCRW,TI,3.5MMX 20MM; BONE FIXATION SCREW/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LO-PRO SCRW,TI,3.5MMX 20MM; BONE FIXATION SCREW/PLATE Back to Search Results
Model Number LO-PRO SCRW,TI,3.5MMX 20MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2025, a sales representative reported via (b)(4) that an ar-8935-20 3.5 mm x 20 mm titanium low-profile screw head broke off cleanly during use.No patient was affected.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a manufacturing issue; however, due to the device not being returned, we are unable to confirm the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LO-PRO SCRW,TI,3.5MMX 20MM
Common Device Name
BONE FIXATION SCREW/PLATE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22542268
Report Number1220246-2025-02960
Device Sequence Number15572556
Product Code HWC
UDI-Device Identifier00888867050464
UDI-Public(01)00888867050464(11)230920(10)15143926
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K203294
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLO-PRO SCRW,TI,3.5MMX 20MM
Device Catalogue NumberAR-8935-20
Device Lot Number15143926
Was Device Available for Evaluation? No
Distributor Facility Aware Date06/25/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2025
Supplement Date Received by Manufacturer09/26/2025
Initial Report FDA Received Date07/18/2025
Supplement Report FDA Received Date10/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-