| Model Number |
3300 |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Correction to section d, suspect medical device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a left ventricular (lv) lead threshold test, this programmer would initiate the threshold test and would not be responsive to the commands.Technical services was consulted and advised on performing a live session to further investigate this behavior.No adverse patient effects were reported.This programmer remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete udi and other product specific information.If additional details become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a left ventricular (lv) lead threshold test, this programmer would initiate the threshold test and would not be responsive to the commands.Technical services was consulted and advised on performing a live session to further investigate this behavior.No adverse patient effects were reported.This programmer remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to section d, suspect medical device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a left ventricular (lv) lead threshold test, this programmer would initiate the threshold test and would not be responsive to the commands.Technical services was consulted and advised on performing a live session to further investigate this behavior.No adverse patient effects were reported.This programmer remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that during a left ventricular (lv) lead threshold test, this programmer would initiate the threshold test and would not be responsive to the commands.Technical services was consulted and advised on performing a live session to further investigate this behavior.No adverse patient effects were reported.This programmer remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for product analysis.Boston scientific performed an investigation with the available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch, and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.Investigation determined that the most common cause of the touchscreen becoming unresponsive during use was the water detection feature within the programmer's liquid crystal display (lcd) touchscreen assembly.The purpose of this water detection feature is to detect the presence of water on/near the programmer screen.If water is detected, this feature is designed to disable the touchscreen to prevent water droplets from causing false positive touch inputs during use.Specifically, the investigation determined that various inputs could induce false positive detections of this water detection feature, which then results in an unresponsive touchscreen.Boston scientific's investigation also identified additional potential causes independent of the water detection feature.These include a variety of unrelated root causes such as hardware, environmental, software, and user-related causes that can be intermittent and not reproducible during laboratory analysis.Correction to section d, suspect medical device.
|
| |
|