• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND HUMERAL HEAD TRIAL 48.5X17; SHOULDER INSTRUMENT - HEAD TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND HUMERAL HEAD TRIAL 48.5X17; SHOULDER INSTRUMENT - HEAD TRIAL Back to Search Results
Catalog Number 620048170
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: h3, h6: a product investigation was completed: visual analysis of the returned device revealed that humeral head trial had four broken internal tabs.The broken fragments were not returned for evaluation.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces while disassembling.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended user error.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
It was reported that offset trial adapter and humeral head trial were discovered broken into multiple pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMERAL HEAD TRIAL 48.5X17
Common Device Name
SHOULDER INSTRUMENT - HEAD TRIAL
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22544665
Report Number1818910-2025-12489
Device Sequence Number12506318
Product Code HWT
UDI-Device Identifier10603295539230
UDI-Public(01)10603295539230(10)321075
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number620048170
Device Lot Number321075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2025
Initial Report FDA Received Date07/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight82 KG
-
-