SYNTHES GMBH EXTRACT-BOLT F/SCR Ø4.5+5; EXTRACTOR Back to Search Results
Catalog Number 309.490
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Necrosis (1971)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the patient underwent an unknown surgery with fns implants.After the surgery, osteonecrosis of the femoral head occurred, and the revision surgery was performed on (b)(6) 2025.(b)(4) is related to (b)(4) which reports the breakage of the jj product occurred at the revision surgery, and (b)(4) which reports the breakage of the polo product occurred at the revision surgery.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found nothing indicative of a device nonconformance.Signs of usage and normal wear, which could have been a result of implantation and explantation.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the extract-bolt f/scr ø4.5+5 would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot part:309.490 lot:4l45596 manufacturing site: werk balsthal supplier: na release to warehouse date: 26 apr 2019 expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9 h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACT-BOLT F/SCR Ø4.5+5
Common Device Name
EXTRACTOR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BALSTHAL (CH)
dornacherstrasse 20
balsthal CO
SZ  
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22545881
Report Number8030965-2025-07285
Device Sequence Number18296480
Product Code HWB
UDI-Device Identifier10886982185761
UDI-Public(01)10886982185761(10)4L45596
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number309.490
Device Lot Number4L45596
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/02/2025
Supplement Date Received by Manufacturer07/23/2025
07/30/2025
09/16/2025
Initial Report FDA Received Date07/18/2025
Supplement Report FDA Received Date07/24/2025
07/31/2025
09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ANTIROTSCR F/FEM NECK SYST F/CONSTRUCT L; BOLT F/FEM NECK SYST F/CONSTRUCT LENGTH; LOCKSCR Ø3.5 SELF-TAP L34 TAN; PL 1-HO F/FEM NECK SYST TAN
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-