|
A health care professional reported that during the cataract surgery with intraocular lens (iol) implant procedure the lens did not load correctly.When the surgeon was about to inject it into the eye he spotted that the leading haptic was protruding forward, and not folded over the top of the lens.The main issue was that the rounded tip of the haptic was missing, rendering the lens unusable.The injector was in contact with the patient, about to be inserted into the incision in the cornea.Once the fault was identified the injector was immediately withdrawn and lens was replaced with another lens, which loaded perfectly.The procedure was completed without further incident.There was no impact to patient.The missing tip of the haptic was definitely not in the eye of the patient, the surgeon had not begun to inject it, he spotted the fault as soon as it came into the field of vision of the microscope, as he was about to offer it up to the eye.The surgeon could see it in the injector.It was also not found in the packaging that the lens was in.There was no resistance or anything to indicate that there was a problem while the scrub practitioner was loading the lens.Additional information has been requested.
|
|
Additional information provided in d.9., h.3., h.6.And h.11.The device was returned in the blister tray.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted back to mid-nozzle.No damage or abnormalities observed.The lens was returned separated from the device, taped to the blister tray.Viscoelastic was dried on the lens.One haptic distal tip was broken (not returned).Both haptics were in a relaxed, original positions.The reported straight leading haptic could not be confirmed.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A qualified viscoelastic was indicated.The reported broken haptic was observed.The root cause cannot be determined.The plunger was retracted.The plunger position in relation to the broken haptic during advancement cannot be determined.The haptic position may indicate the plunger was not fully advanced.If the plunger is not fully advanced the trailing haptic may not release properly from the device.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the instructions for use (ifu).A qualified viscoelastic was indicated.The ifu instructs: take at least 7 seconds to gently fold the intraocular lens (iol) by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the ¿fill-to¿ line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.After the lens has been advanced to the ¿fill-to¿ line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.No part of the lens should exit the nozzle prior to insertion through the incision.Once the lens is in position at the ¿fill-to¿ line, it should be implanted within 3 minutes.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome.Broken haptics may occur: ¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.In addition, haptic strength (modulus) decreases as the temperature increases and is more likely to break under stress.¿ if a straight trailing haptic occurs and it was not properly detected to be out of position.¿ if the plunger is not fully advanced, or if the plunger is allowed to retract, the trailing haptic may not release properly from the device.Any of the above listed causes alone, or in combination, may create the reported event.The ifu instructs: insert the tip of the cartridge through the incision.Position the tip at the anterior capsule opening.If the leading haptic is looped or straight in front of the optic, rotate the cartridge clockwise as needed to be bevel left to facilitate placement of the leading haptic into its normal orientation in the bag.Slowly advance the lens until the optic is exiting the nozzle.Rotate the cartridge counter-clockwise towards bevel right as needed to place the lens anterior side up.A straight leading haptic may occur: due to the normal folding variations as indicated the ifu diagrams.If the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.If the ovd fill rate is too fast the folding effect on the leading haptic is minimized.If there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|