| Model Number |
470344-14 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the 8 mm curved bipolar dissector instrument to perform failure analysis.Failure analysis investigations confirmed the customer reported complaint.The instrument was analyzed and it was found to have a dislodged conductor wire.The wire was sticking out from the distal clevis and a loop was protruding above the outer surface of main tube.The wire insulation was not damaged and the instrument passed an electrical continuity test.The probable root cause of dislodged conductor wire is attributed to component susceptibility to damage through external collisions, during use, or during reprocessing.Based on additional information, it has been determined that the instrument involved with this complaint does not meet the criteria for isifa2024-09-c as previously listed in section h9.
|
| |
|